Error: sample fluidics high pressure limit (value) on ACQUITY FTN - WKB9194

Article number: 9194
This article is a portion of a Troubleshooting Path.  To begin troubleshooting this issue, please select the link below.
Fix Sample Fluidics High Pressure Limit Error
A troubleshooting path for the Sample Fluidics High Pressure Limit Error
Pages: 5

SYMPTOMS

  • Sample Fluidics High Pressure Limit (pressure value)
  • Flush path with fittings 2 and 3 loose, prime syringe, but error is still present

ENVIRONMENT

  • ACQUITY H-Class FTN
  • ACQUITY I-Class FTN

CAUSE

Error occurs when the pressure exceeds 210 psi in the low-pressure fluidics path.

FIX or WORKAROUND

Replace needle and needle seal. 

ADDITIONAL INFORMATION

 

overpressure

Not able to find a solution? Click here to request help.