UV/Vis Absorbance Detector - Lower response than expected - WKB89944

Article number: 89944

SYMPTOMS

  • Lower-than-expected peak responses when compared to other similar systems.
  • Retention times are ok.
  • Peak Areas reproducible.
  • Peak Heights reproducible.

ENVIRONMENT

  • 2487 Dual-Wavelength Absorbance Detector
  • 2489 UV/Vis Detector
  • 2996 PDA Detector
  • 2998 Photodiode Array Detector
  • ACQUITY UPLC Tunable UV Detector
  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC PDA eλ Detector

CAUSE

The detector has not been powered-off for a significant length of time, so the calibration against the internal filters has drifted over time. When you select a wavelength for a method, the detector is actually reporting back from the side of a calibration peak rather than the top.

FIX or WORKAROUND

Power cycle the Detector and allow the Detector to reinitialize and perform internal self-adjustments.

ADDITIONAL INFORMATION