No peaks observed, pump B primary is flat-lining with ACQUITY I-Class - WKB89796

Article number: 89796

SYMPTOMS

  • No peaks observed
  • Flow lacking or pulsating from pump B vent line when priming
  • Pump B primary fails to compress when performing dynamic leak test

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS

CAUSE

Primary check valve cartridge failed. 

FIX or WORKAROUND

Replace the the primary check valve cartridge (700004139) and perform the dynamic leak test.

ADDITIONAL INFORMATION

 

id89796, checkvalve, DLT, UPBINARY, UPBSM+, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM