Alliance SQT 2489/2998 detector linearity and height sensitivity failure - WKB89586
Article number: 89586
SYMPTOMS
- PM performed; all diagnostic leak tests pass
- Passing injector linearity and wavelength accuracy test
- Detector linearity test failure
- Detector linearity response checked; peak height outside detector's linear range as measured 3.1 AU with 0.22 mg/mL concentration
- Recommended detector peak heights:
- < 2.0 AU for 2996/2998
- < 2.5 AU for the 2487/2489
- Recommended detector peak heights:
ENVIRONMENT
- Alliance 2690/2695 Separations Module
- Alliance - SQT part 667003309 RevN
- 2487/2489
- 2996/2998
CAUSE
Alliance system incorrectly configured for 100-µL syringe
FIX or WORKAROUND
Configure 250-µL syringe.
ADDITIONAL INFORMATION