No pressure on ACQUITY UPLC H-Class in inject mode - WKB8953

Article number: 8953

SYMPTOMS

  • No pressure when FTN sample manager is in inject position
  • Normal pressure when in load position
  • Solvent manager (QSM) passes leak tests

ENVIRONMENT

  • ACQUITY UPLC H-Class

CAUSE

Faulty needle seal

FIX or WORKAROUND

Replace the needle seal and seat port as described in ACQUITY UPLC SM-FTN Overview and Maintenance Information.