Skip to main content
Waters

Alliance sample syringe overpressure error during system prep or prime - WKB89219

Article number: 89219

SYMPTOMS

  • System flows normally during regular operation
  • Purge injector completes without error
  • Error occurs during wet prime, showing pressure of >350 psi and stopping flow

ENVIRONMENT

  • Alliance HPLC System

CAUSE

Waste (clear colored) line out of V3 is kinked or blocked.

V3 (700001229) has failed.

FIX or WORKAROUND

  1. Remove blockage from the V3 waste line.
  2. If blockage is not the issue, replace V3.

ADDITIONAL INFORMATION

 

id89219, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB

Not able to find a solution? Click here to request help.