ACQUITY FLR lamp lighting failure - WKB89162
Article number: 89162
SYMPTOMS
- Lamp lighting failure during diagnostics
- Reseated lamp and issue persists
ENVIRONMENT
ACQUITY UPLC FLR Detector
CAUSE
Faulty lamp.
FIX or WORKAROUND
Replace lamp part number 201000193
ADDITIONAL INFORMATION
New lamp requires wavelength calibration and normalization. Signal to noise should be >1000.