"Lost prime" error during run with 2695 module - WKB88680
Article number: 88680
SYMPTOMS
- "Lost prime" error during run
- Run aborts
ENVIRONMENT
- 2695 Separations Module
- Non-Empower software (Agilent or Thermo)
CAUSE
Degasser is off in the instrument method.
FIX or WORKAROUND
- Turn on the degasser in the instrument method.
- Verify resolution of the "lost prime" error.
ADDITIONAL INFORMATION