The mobile phase/solvent in the bottle runs out or decrease quickly even though it’s not flowing with H-Class QSM - WKB88061
Article number: 88061
SYMPTOMS
- The mobile phase/solvent in the bottle runs out or decreases quickly even though it’s not flowing
ENVIRONMENT
- ACQUITY UPLC H-Class
CAUSE
GPV (gradient proportioning valve) inlet/outlet tubing is loose and leaking
FIX or WORKAROUND
- Tighten the screws of each tube on the gradient profiling valve (the 15 in the figure below).

ADDITIONAL INFORMATION
If the leak is coming from the drip tray indicating the leak is internal, replace the GPV (pt # 700005161) for the line loosing mobile phase.
