The mobile phase/solvent in the bottle runs out or decrease quickly even though it’s not flowing with H-Class QSM - WKB88061

Article number: 88061

SYMPTOMS

  • The mobile phase/solvent in the bottle runs out or decreases quickly even though it’s not flowing

ENVIRONMENT

  • ACQUITY UPLC H-Class

CAUSE

GPV (gradient proportioning valve) inlet/outlet tubing is loose and leaking

FIX or WORKAROUND

  1. Tighten the screws of each tube on the gradient profiling valve (the 15 in the figure below).
    clipboard_e6e4c6739d0e0a8df5fdfc44b0495e4f7.png

ADDITIONAL INFORMATION

If the leak is coming from the drip tray indicating the leak is internal, replace the GPV (pt # 700005161) for the line loosing mobile phase.