  • If a gradient table contains a rapid change from 100%A to 100%B (or vice versa), the Mixer can clog, especially if it is older and already has blockages forming.
  • Pressure issues occur gradually but eventually manifest themselves in a full overpressure.
  • Pressure traces show jagged tops, demonstrating the resistance to smooth fluidics.