High pressure on an ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM) - WKB87393

Article number: 87393

SYMPTOMS

  • High pressure on an ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • Cleaning with warm water does not resolve the issue
  • Different columns do not resolve the issue

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

The system is clogged.

FIX or WORKAROUND

 

Remove column and install an Union.

Flush with 5% formic acid with MeOH: IPA: ACN: H2O.

Flush with H2O until neutralization of pH.