Error "Needle Seal residual force exceeded" on startup of ACQUITY UPLC Sample Manager FTN - WKB87274
Article number: 87274
SYMPTOMS
- "Needle Seal residual force exceeded" error upon startup
- Resetting in the console does not clear the error
- Power cycling does not resolve the error
ENVIRONMENT
- ACQUITY H-Class Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY I-Class Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY H-Class Sample Manager - Flow Through Needle Plus (SM-FTN Plus)
- ACQUITY I-Class Sample Manager - Flow Through Needle Plus (SM-FTN Plus)
- ACQUITY H-Class Sample Manager - Flow Through Needle Bio (SM-FTN Bio)
- ACQUITY H-Class Sample Manager - Flow Through Needle Bio Plus (SM-FTN Bio Plus)
- ACQUITY APC Sample Manager - Flow Through Needle (SM-FTN)
CAUSE
Needle is bent or has come out of the needle holder.
FIX or WORKAROUND
- If possible, place the needle back in the holder.
- Replace the needle.
- Reset the sample manager through the console.
ADDITIONAL INFORMATION