2998 lamp is red and error message:Wavelength calibration failure - WKB86910
Article number: 86910
SYMPTOMS
- Lamp led is red.
- Wavelength calibration failure.
- Replacing lamp does not resolve issue.
- Cleaning cell with organic solvent does not resolve issue.
ENVIRONMENT
2998 Photodiode Array Detector
CAUSE
One or more calibration peak accuracy not <1 nm.
FIX or WORKAROUND
- Perform an erbium calibration (see WKB6197).
- Save new calibration values Lambda 1 and Lambda 512.
- Reset 2998.
- Perform wavelength verification.
ADDITIONAL INFORMATION