ACQUITY UPLC Sample Manager - Flow Through Needle fails needle seal readiness test - WKB86641
Article number: 86641
SYMPTOMS
- ACQUITY UPLC Sample Manager–FTN fails needle seal readiness test with negative pressure difference result
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
CAUSE
A 0.06-µL/min LC flow rate is too low with 150-cm column selection, even though pressure is stable at 5000 psi.
FIX or WORKAROUND
Change to standard ACQUITY UPLC BEH C18 1.7µm 2.1x50mm column (186002350). (A 0.4-µL/min flow passes with 5000 psi.)
ADDITIONAL INFORMATION