ACQUITY UPLC Sample Manager - Flow Through Needle fails needle seal readiness test - WKB86641

Article number: 86641

SYMPTOMS

  • ACQUITY UPLC Sample Manager–FTN fails needle seal readiness test with negative pressure difference result        

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class

CAUSE

A 0.06-µL/min LC flow rate is too low with 150-cm column selection, even though pressure is stable at 5000 psi.

FIX or WORKAROUND

Change to standard ACQUITY UPLC BEH C18 1.7µm 2.1x50mm column (186002350). (A 0.4-µL/min flow passes with 5000 psi.)

ADDITIONAL INFORMATION

 