Intermittent overpressure during injection with 1525 Micro and 2777 - WKB86438
Article number: 86438
SYMPTOMS
- System gives overpressure error after an injection
ENVIRONMENT
- 1525 Micro Binary HPLC Pump
- 2777 Sample Manager
CAUSE
Particles are introduced into the injector valve during an injection, which causes the valve to block.
The most common cause of particle introduction is use of foil to cover sample plates. A better solution would be to use pre-slit PTFE/silica plate mats.
FIX or WORKAROUND
- Flush all the valve ports with the LC pump while switching the valve position to clear the blockage.
- If the rotor seal has been damaged, this will need to be replaced as well.
ADDITIONAL INFORMATION
id86438, 1525, 1525U, 1525UIVD, 2777, 2777C, 2777IVD