Error on Alliance 2695: sample transducer range (4.0) - WKB859

Article number: 859

SYMPTOMS

  • "Sample transducer range (4.0)"
  • 8800 psi

ENVIRONMENT

  • Alliance Separations Module 2695

CAUSE

Loose sample transducer connection or bad transducer cable

FIX or WORKAROUND

 

 Replace sample transducer cable - 2690 PRES TRANSDUCER CABLE, part number WAT270906.

ADDITIONAL INFORMATION

With the transducer disconnected, a "sample transducer range (4.0)" error occurs during initialization. In Diagnostics, the pressure is 8800 psi; in Service Utilities, the transducer voltage is 0 volts.

 