What are SQT manual OQ temperature limits on 30-cm Column Heater? - WKB85679

Article number: 85679

ENVIRONMENT

  • Alliance
  • SQT

ANSWER

Update on 13May2025: 

Alliance SQT Revision D (PN: 667006605DDRevD), November 2024.

clipboard_ef8a6bc47f67d29fef037e6f2ac0963e2.png

This is the details of temperature related from release note of 715007961v02: SystemsQT for Waters Alliance HPLC Systems Change Note.

clipboard_e118df68f034121e6637284924568b534.png

 

Original:

Measured at the temperature sensor, the temperature accuracy of the sample heater/cooler is ±3.0 °C. To compensate for typical measurement inaccuracy, the temperature accuracy specification appearing in the temperature accuracy report gives an allowable temperature variation of ±4 °C.

 