2998 baseline ripple with Prep 150 system - WKB85462

Article number: 85462

SYMPTOMS

  • Baseline ripple at analytical run with new Prep 150 system, but ripple is not visible in prep run
    clipboard_e7489f1a5c68bd2a49c1671a9be83b193.png
    clipboard_eb50c9bb66b718123ba44b7ab5a9df933.png

ENVIRONMENT

  • ChromScope
  • 2998 Photodiode Array Detector
  • Prep 150 LC System
  • Manual injector

CAUSE

Lack of back pressure at PDA

FIX or WORKAROUND

Add 5 to 6 inches of 0.010-inch ID PEEK tubing (430001945) to the outlet of the 2998 autopure flow cell before going to WFCIII.

ADDITIONAL INFORMATION


id85462, 2998, 2998B, 2998SFC, A-10PDA, baseline noise