Column Manager column outlet tubing leaks- WKB84738
Article number: 84738
SYMPTOMS
- Column outlet tubing leaks
ENVIRONMENT
- ACQUITY UPLC Column Manager
CAUSE
Broken column outlet tubing
FIX or WORKAROUND
Replace tubing and check all fluidic connections are correctly tightened up.
H-class tubing 700005478 sold as Welded Tube Assembly, SS, 19.0 LG, LP
I-Class tubing 700008868 sold as Welded Tube Assy, SS, .003 ID X 19.0 LG
ADDITIONAL INFORMATION
If you have a BIO system check this link : What are the part numbers for Bio Column Manager tubing? - WKB196996
For I-Class PLUS see What is the part number for the ACQUITY UPLC I-Class PLUS column outlet tubing to CM-A valve? - wkb225963 - Waters