Alliance 2695 Separations Module flow rate is low - WKB83058
Article number: 83058
SYMPTOMS
- Flow rate is low
- Passes compression check but fails static leak test
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
The seal wash line is dry.
FIX or WORKAROUND
- Manually prime the seal wash line until the line is full.
- Prime solvent lines.
ADDITIONAL INFORMATION