Seal wash does not prime correctly on H-Class QSM - WKB83011
Article number: 83011
ACQUITY UPLC H-Class
Quaternary Solvent Manager (QSM)
SYMPTOMS
- Seal wash does not prime correctly
- Solvent filter is new
ENVIRONMENT
ACQUITY UPLC H-Class
Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC H-Class
-
ACQUITY UPLC H-ClassQuaternary Solvent Manager (QSM)
CAUSE
Improper priming.
FIX or WORKAROUND
- Disconnect the seal wash tube outlet from the drip tray.
- Introduce a 10-mL plastic syringe at the end of the tube.
- Launch the "Prime Seal Wash" function from the console menu.
- Aspirate with the syringe until seal wash solvent arrives at the seal wash pump.
- Leave the "Prime Seal Wash" function running for a few minutes.
- Stop the "Prime Seal Wash" function.
- Reconnect the seal wash tube to the drip tray.