ACQUITY UPLC Sample Manager: SQT Injector Linearity Failed - WKB8292
Article number: 8292
SYMPTOMS
-
Injector linearity failed with lower than expected R2 Values. Injection Accuracy Pass.
-
System Precision test passes with 0.3% area RSD.
-
-
Maintenance performed.
-
System primed thoroughly, no air bubbles.
-
All leak test diagnostics pass. Air sensor calibration successful.
-
Characterization of the needle and loop volumes are repeatable with reproducible results.
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY SystemsQT
CAUSE
Poor sample loop fitting installation creating a dead volume at inject valve port positions 1 and 4.
FIX or WORKAROUND
Replace sample loop, 430001326 for the LOOP, SAMPLE, 10 uL, STANDARD.
ADDITIONAL INFORMATION
Following the sample loop replacement, system precision additionally improved to 0.1% area RSD.