Degasser vacuum pump over pressure - WKB82568

Article number: 82568

SYMPTOMS

  • Turning on the ACQUITY UPLC BSM gives the error "Degasser Vacuum Pump Over Pressure"
  • Power cycled the ACQUITY UPLC BSM several times, but same error occurs
  • Other errors that appear are degasser pressure not reached 1.3, degasser no vacuum and degasser electronics failure

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY UPLC QSM 

CAUSE

The lines are dry or have air in them

FIX or WORKAROUND

  1. Power-on the ACQUITY UPLC BSM / QSM.
  2. Prime all solvent lines thoroughly.