NanoACQUITY or M-Class System overpressure - WKB8168
Article number: 8168
SYMPTOMS
- H/W overpressure error for the nano Binary Solvent Manager (nBSM).
- Start flow and it quickly spikes to a high pressure.
ENVIRONMENT
- NanoACQUITY
- ACQUITY M-Class
CAUSE
Debris blocking the lines. Potentially contaminated or low quality solvents.
FIX or WORKAROUND
- Prepare fresh mobile phases with new, unopened LCMS grade solvents and additives.
- Flush with high organic in Analytical mode overnight at as high a flow rate as possible without getting overpressure.
- If there is still high pressure, disconnect the trap column from port 6 of the inject valve and plumb some PEEK tubing to port 6 going to waste (or place a wad of Kim wipes below the inject valve to catch the solvent).
- Prime A and B, and then flush with acetonitrile/water/methanol/isopropyl alcohol and 0.2% formic acid for at least three hours at 15 uL/min.
- Prime and flush with fresh normal A and B solvents (water for A and acetonitrile for B) for 15 minutes at 15 uL/min.
- Stop flow and attach the trap column.