2424 Poor area and height reproducibility - WKB78751
Article number: 78751
SYMPTOMS
- Poor area and height reproducibility
- The injection accuracy and reproducibility of 2695 is accurate and reproducible .
- The baseline noise of 2424 is normal
- The nebulizer spray is normal
- The siphon tube is installed correctly
ENVIRONMENT
- 2695 Separations Module
- 2424 Evaporative Light Scattering Detector
CAUSE
The drift tube has been polluted (contaminated), the dirty material may absorb the sample,or affect drift tube temperature .
FIX or WORKAROUND
- Set nebulizer heater level to 100%,drift tube temperature to 100 degrees and gas pressure to 35 Psi.
- Replace the column with a union
- Set flow to 1 ml/min ,flush the system for 4 hours with solvent of Methanol:Acetonitrile:water: Isopropanol =1:1:1:1 added 2% formic acid
- Set flow to 1 ml/min ,flush the system for 4 hours with solvent of Acetone:water=1:1
ADDITIONAL INFORMATION
id78751, 2420, 2424, 2424B, baseline ripple