2424 Poor area and height reproducibility - WKB78751

Article number: 78751

SYMPTOMS

  • Poor area and height reproducibility
  • The injection accuracy and  reproducibility of 2695 is accurate and reproducible .
  • The baseline noise of 2424 is normal
  • The nebulizer spray is normal
  • The siphon tube is installed correctly

ENVIRONMENT

  • 2695 Separations Module
  • 2424 Evaporative Light Scattering Detector

CAUSE

The drift tube has been polluted (contaminated), the dirty material may absorb the sample,or affect drift tube temperature .

 

FIX or WORKAROUND

  1. Set nebulizer heater level to 100%,drift tube temperature to 100 degrees and gas pressure to 35 Psi.
  2. Replace the column with a union
  3. Set flow to 1 ml/min ,flush the system for 4 hours with solvent of Methanol:Acetonitrile:water: Isopropanol =1:1:1:1 added 2% formic acid
  4. Set flow to 1 ml/min ,flush the system for 4 hours with solvent of Acetone:water=1:1

ADDITIONAL INFORMATION

 

id78751, 2420, 2424, 2424B, baseline ripple