Probe temperature failed to settle error message on ACQUITY QDa Mass Detector - WKB78072
Article number: 78072
SYMPTOMS
- Probe temperature failed to settle error message.
- Found: Stop flow input activated message appears in logs just prior to probe temperature error.
- Found: Tank pressure for N2 supply is less than 100 PSI.
ENVIRONMENT
ACQUITY QDa Mass Detector
CAUSE
Tank pressure for N2 supply is less than the require pressure to the back of the instrument.
FIX or WORKAROUND
- Replace N2 supply tank.
ADDITIONAL INFORMATION
id78072, QDA