2424 ELS detector signal decreases through a sample set - WKB77718

Article number: 77718

SYMPTOMS

  • The signal decreases through a sample set and becomes noise
  • Method runs successfully on a different, non- ELS detector
  • No liquid comes from the waste tube at normal flow of 1 mL/min, even after adjusting the drain tubing

ENVIRONMENT

  • 2424 Evaporative Light Scattering Detector
  • ACQUITY UPLC ELS Detector

CAUSE

Dirty nebulizer and drift tube

FIX or WORKAROUND

Clean the nebulizer and drift tube per the instructions in the model of detector you have

 See : 2424 Operator's Guide 

See : ACQUITY UPLC Evaporative Light Scattering Detector Getting Started Guide

ADDITIONAL INFORMATION

 

id77718, 2424, 2424B, UPELS, UPELSARC