2424 ELS detector signal decreases through a sample set - WKB77718
Article number: 77718
SYMPTOMS
- The signal decreases through a sample set and becomes noise
- Method runs successfully on a different, non- ELS detector
- No liquid comes from the waste tube at normal flow of 1 mL/min, even after adjusting the drain tubing
ENVIRONMENT
- 2424 Evaporative Light Scattering Detector
- ACQUITY UPLC ELS Detector
CAUSE
Dirty nebulizer and drift tube
FIX or WORKAROUND
Clean the nebulizer and drift tube per the instructions in the model of detector you have
See : 2424 Operator's Guide
See : ACQUITY UPLC Evaporative Light Scattering Detector Getting Started Guide