Skip to main content
Waters

2424 ELS detector signal decreases through a sample set - WKB77718

Article number: 77718

SYMPTOMS

  • The signal decreases through a sample set and becomes noise
  • Method runs successfully on a different, non- ELS detector
  • No liquid comes from the waste tube at normal flow of 1 mL/min, even after adjusting the drain tubing

ENVIRONMENT

  • 2424 Evaporative Light Scattering Detector
  • ACQUITY UPLC ELS Detector

CAUSE

Dirty nebulizer and drift tube

FIX or WORKAROUND

Clean the nebulizer and drift tube per the instructions in the model of detector you have

 See : 2424 Operator's Guide

See : ACQUITY UPLC Evaporative Light Scattering Detector Getting Started Guide

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.