2489 Detector does not collect data after inject start - WKB77349
Article number: 77349
SYMPTOMS
- 2489 Detector is not collecting data after inject start
- During monitoring, column conditioning, or equilibration, the detector collects data
ENVIRONMENT
- 2489 UV/Vis Detector
CAUSE
The Inject Start event is set to "Ignore" on the 2489.
FIX or WORKAROUND
Modify Inject event in: "Low", using the second page in the Configuration screen
ADDITIONAL INFORMATION
id77349, 2489, 2489B, 2489SFC, A-10UV