2489 Detector does not collect data after inject start - WKB77349

Article number: 77349

SYMPTOMS

  • 2489 Detector is not collecting data after inject start
  • During monitoring, column conditioning, or equilibration, the detector collects data

ENVIRONMENT

  • 2489 UV/Vis Detector

CAUSE

The Inject Start event is set to "Ignore" on the 2489.

FIX or WORKAROUND

Modify Inject event in: "Low", using the second page in the Configuration screen

ADDITIONAL INFORMATION

 

id77349, 2489, 2489B, 2489SFC, A-10UV