ACQUITY UPLC BSM / QSM gives overpressure error - WKB76470

Article number: 76470

SYMPTOMS

  • System gives "pump h/w overpressure" error
  • Bypassing the column does not show excess pressure

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY ARC
  • ACQUITY ARC Bio

CAUSE

The column is clogged or defective.

FIX or WORKAROUND

  1. Clean / Flush Column according to Care and Use manual
  2. Replace the column