ACQUITY UPLC BSM / QSM gives overpressure error - WKB76470
Article number: 76470
SYMPTOMS
- System gives "pump h/w overpressure" error
 - Bypassing the column does not show excess pressure
 
ENVIRONMENT
- ACQUITY UPLC
 - ACQUITY UPLC BSM
 - ACQUITY UPLC H-Class
 - ACQUITY UPLC H-Class Bio
 - ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
 - ACQUITY UPLC I-Class
 - ACQUITY UPLC H-Class PLUS
 - ACQUITY UPLC H-Class PLUS Bio
 - ACQUITY UPLC I-Class PLUS
 - ACQUITY ARC
 - ACQUITY ARC Bio
 
CAUSE
The column is clogged or defective.
FIX or WORKAROUND
- Clean / Flush Column according to Care and Use manual
 - Replace the column
 
ADDITIONAL INFORMATION
id76470, A-30SDM, UPAQSM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBINARY, UPBINARY, UPBSM+, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM, UPPDAARCB, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO
 