"Preparing for injection" message while running a sample set in Empower - WKB76346
Article number: 76346
SYMPTOMS
- Sample set stuck on "Preparing for injection" while performing single injection
- After disabling one detector from the instrument method, "Preparing for injection" message does not go away
ENVIRONMENT
- ACQUITY H-Class
- ACQUITY TUV
- ACQUITY PDA
- ACQUITY Arc
- ACQUITY SQ Detector
- ACQUITY SQ Detector 2
- ACQUITY TQ Detector
- ACQUITY QDa Mass Detector
- ARC HPLC
- Empower 3
CAUSE
The modules included in the chromatographic system do not match the modules used in the instrument method.
FIX or WORKAROUND
- Create a separate chromatographic system within Empower that includes only the modules used in the instrument method.
- If you have a separate chromatographic system, use the system that matches the instrument method.