"Preparing for injection" message while running a sample set in Empower - WKB76346

Article number: 76346

SYMPTOMS

  • Sample set stuck on "Preparing for injection" while performing single injection
  • After disabling one detector from the instrument method, "Preparing for injection" message does not go away

ENVIRONMENT

  • ACQUITY H-Class
  • ACQUITY TUV
  • ACQUITY PDA
  • ACQUITY Arc
  • ACQUITY SQ Detector
  • ACQUITY SQ Detector 2
  • ACQUITY TQ Detector
  • ACQUITY QDa Mass Detector
  • ARC HPLC
  • Empower 3

CAUSE

The modules included in the chromatographic system do not match the modules used in the instrument method.

FIX or WORKAROUND

  1. Create a separate chromatographic system within Empower that includes only the modules used in the instrument method.
  2. If you have a separate chromatographic system, use the system that matches the instrument method.

ADDITIONAL INFORMATION

 