Error (with a 2707 Autosampler): Missing plate or rack - WKB7624
Article number: 7624
SYMPTOMS
- "Missing plate or rack" error message in control panel
- Resetting or rebooting the 2707 does not correct the problem
- Error occurs regardless of plate or vial location
- Remove the plate from the carousal and reseatting doesn't correct problem
ENVIRONMENT
- Empower
- 2707 Autosampler
CAUSE
Damaged needle
FIX or WORKAROUND
- Replace the needle (700003842, NEEDLE ASSY, STD) and recalibrate.
ADDITIONAL INFORMATION
Waters 2707 Autosampler Operator’s Guide