Retention Times Increasing - WKB76187
Article number: 76187
SYMPTOMS
- The retention times of sample and standard peaks increasing
- Static leak test failure
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Worn pump components
FIX or WORKAROUND
Replace the plunger seals, face seals, plunger seal wash seals, and seal wash seals.
ADDITIONAL INFORMATION
700004073 - Alliance Solvent Manager PM kit
Or,
WAT270939 - SEAL, FACE, PKG (4)
WAT271018 - PLUNGER, SEAL WASH, PKG (2), 2690
WAT270938 - SEAL, PLUNGER, PKG (2)
2 x WAT271017 - SEAL, FACE, SEALWASH, PKG (1)