Console shows error : Microsoft.NET Framework error: Unhandled exception has occurred in your application - WKB75810

Article number: 75810

SYMPTOMS

  • Microsoft .net framework error occurs when opening the console. Unhandled exception has occurred in your application.
  • Details information, System.Exception: An InstrumentImage requires a name so that the position can be saved correctly.

2432 unhandled exception error.JPG

ENVIRONMENT

  • 2432 Conductivity Detector
  • ACQUITY UPLC
  • Empower 3 Feature Release 5

CAUSE

ICS or firmware is corrupt

FIX or WORKAROUND

  • Uninstall and reinstall driver and firmware
  • If driver and firmware reinstall doesn't resolve, recreate chromatographic system

 

ADDITIONAL INFORMATION

 