Console shows error : Microsoft.NET Framework error: Unhandled exception has occurred in your application - WKB75810
Article number: 75810
SYMPTOMS
- Microsoft .net framework error occurs when opening the console. Unhandled exception has occurred in your application.
- Details information, System.Exception: An InstrumentImage requires a name so that the position can be saved correctly.
ENVIRONMENT
- 2432 Conductivity Detector
- ACQUITY UPLC
- Empower 3 Feature Release 5
CAUSE
ICS or firmware is corrupt
FIX or WORKAROUND
- Uninstall and reinstall driver and firmware
- If driver and firmware reinstall doesn't resolve, recreate chromatographic system