ACQUITY autosampler tray position error for vial position in Xcalibur with SII driver pack - WKB75442
Article number: 75442
SYMPTOMS
- A * is shown next to position in Xcalibur sample list with "Position is invalid for configured autosampler tray type" error message
- ANSI 48 Vial 2mL Holder is configured autosampler tray
- 1:A,1 plate referencing is used
ENVIRONMENT
- ACQUITY UPLC
- Xcalibur 4.2
- Thermo SII 1.5 driver pack
CAUSE
Tray referencing has changed with the SII driver pack.
FIX or WORKAROUND
- For SII driver pack installations, use 1:A1 vial referencing, minus the comma(,).
- Verify that the * is gone and injection proceeds as expected.
ADDITIONAL INFORMATION
For LC Devices driver pack installations, the traditional 1:A,1 plate referencing is still used.
