High back pressure with a new column on an ACQUITY UPLC I-Class - WKB75330
Article number: 75330
SYMPTOMS
- High back pressure with a new column
- Pressure higher than 10K
ENVIRONMENT
- ACQUITY UPLC I-Class
CAUSE
Unknown
FIX or WORKAROUND
- Loosen the finger-tight fittings on the inject valve at ports 2 and 3.
- Reset the SM-FTN from the console.
- Prime the wash and purge solvents.
- Wash the needle.
- Retighten the fittings.
- Repeat the prime.
- Start flow slowly, increasing to the initial conditions, and allow pressure to stabilize.
- Characterize the needle seal.
ADDITIONAL INFORMATION
id75330, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+