Alliance 2795 Separations Module system overpressure error - WKB74450

Article number: 74450

SYMPTOMS

  • System overpressure error

ENVIRONMENT

  • Alliance HPLC System
  • 2795 Separations Module

CAUSE

The injection valve to column switching valve/column tubing is blocked.

FIX or WORKAROUND

Replace the tubing.

ADDITIONAL INFORMATION

The valve-to-column tubing is often replaced with 0.005" ID PEEK tubing. Over time, the Rheodyne metal ferrules impinge the tubing, causing a blockage. 