2475 configuration not found - WKB7435
Article number: 7435
SYMPTOMS
- When the 2475 detector is turned on, "configuration not found" and "reset to default" error messages appear.
- Once the detector warms, "method not found" and "PMT not calibrated" error messages appear.
ENVIRONMENT
- 2475 Fluorescence (FLR) Detector
CAUSE
3v battery is dead.
FIX or WORKAROUND
- Replace the yellow Snaphat 3v battery (700001834) located internally on the CPU 2000 PCB board.
- Calibrate PMT. See article 2694.
ADDITIONAL INFORMATION