ACQUITY APC shows poor peak height reproducibility - WKB73880
Article number: 73880
SYMPTOMS
- Variable peak heights
- Sample syringe leak test failure
ENVIRONMENT
- ACQUITY APC
- ACQUITY APC Sample Manager - FTN
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The purge solvent ran dry, and insufficient prime cycles were performed; air is still present in tubing.
FIX or WORKAROUND
- Prime the syringe using 30 cycles.
- Perform the sample syringe leak test.
ADDITIONAL INFORMATION
30-50 syringe cycles are recommended if purge solvent runs dry.
id73880, A-30SM, UPASMFTN, UPBINARY, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+