Skip to main content
Waters

ACQUITY APC shows poor peak height reproducibility - WKB73880

Article number: 73880

SYMPTOMS

  • Variable peak heights
  • Sample syringe leak test failure

ENVIRONMENT

  • ACQUITY APC
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

The purge solvent ran dry, and insufficient prime cycles were performed; air is still present in tubing.

FIX or WORKAROUND

  1. Prime the syringe using 30 cycles.
  2. Perform the sample syringe leak test.

ADDITIONAL INFORMATION

30-50 syringe cycles are recommended if purge solvent runs dry.

Not able to find a solution? Click here to request help.