Solvent streams from column inlet tubing during wet prime on Alliance system - WKB73582

Article number: 73582

SYMPTOMS

  • During prime and purge, there is a steady stream of solvent coming from the column inlet tubing
  • New seal pack seals

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Failure of high pressure valve 1 and high pressure valve 2  

FIX or WORKAROUND

  1. Rebuild high pressure valve 1 (WAT045424) and high pressure valve 2 (WAT045424).
  2. Prime and purge with 100% methanol to ensure that issue is resolved.
  3. Run compression check.