Fraction Manager Analytical error "Needle positions have not been taught" appears in the Console - WKB72350

Article number: 72350

SYMPTOMS

  • "Needle positions have not been taught" observed in console.

ENVIRONMENT

  • ACQUITY Console
  • Fraction Manager Analytical
  • Zp axis calibrated successfully

CAUSE

The vessel offset values are incorrect.

FIX or WORKAROUND

Calibrate the vessel offset values and save.

ADDITIONAL INFORMATION

 