Baseline oscillates with 2414 RI detector - WKB72244

Article number: 72244

SYMPTOMS

  • Baseline oscillates

2414 Cycling baseline 26JUL2019.jpg

ENVIRONMENT

  • 2414 Refractive Index Detector
  • 2410 Differential Refractometer
  • 2695 Separations Module
  • Solvents recently switched

CAUSE

The degasser is set to "OFF".

FIX or WORKAROUND

  1. Turn the degasser on.
  2. Purge the detector with degassed solvent.
  3. End purge and allow the system to equilibrate to a stable baseline.