717 Autosampler - Leak inside sample compartment - WKB71460

Article number: 71460

SYMPTOMS

  • Leak inside sample compartment
  • Pressure lower than usual.
  • Loss of mobile phase.

ENVIRONMENT

717plus Autosampler

CAUSE

  1. Sample loop leaking (split loop).
  2. loose or over tightened compression fitting.

FIX or WORKAROUND

  1. Replace sample loop (200 ul: WAT045584/2 ml: WAT069106).
  2. Tighten compression fittings, if still leaking replace them.
  3. Perform injector purge and compression check.

ADDITIONAL INFORMATION