Skip to main content
Waters

Negative drops in baseline and random pressure fluctuations during GPC with 2695 Alliance and RI - WKB70936

Article number: 70936

SYMPTOMS

  • Negative drops in baseline
  • During run, random pressure fluctuations that do not correspond to negative drops
  • Recent system PM
  • Degasser chambers and pump recently replaced

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2410 Differential Refractometer
  • 2414 Refractive Index Detector

CAUSE

Partial clog in tubing from dirty solvents

FIX or WORKAROUND

  1. Replace tubing from column to detector and any tubing joining columns together. 
  2. Inspect solvent bottles and make sure that there are no visible particulates in the bottles.
  3. If there are particulates, replace or thoroughly clean the bottles. 
  4. Replace solvent filters.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.