Skip to main content
Waters

Arc cycling baseline during gradient - WKB69452

Article number: 69452

SYMPTOMS

  • Sinusoidal baseline during gradient
  • Magnitude of wave changes depending on path
  • Magnitude changes as gradient progresses
  • Number of cycles per minute in the baseline is similar to number of cycles of pump strokes per minute
  • Delta psi is 15-16 psi, 0.4% of system pressure

    clipboard_e204abb2b9a0c818fc445361c4ce9fa22.png
    clipboard_e5d46e05db578a07e3b53aa32bf3f5afb.png

     

ENVIRONMENT

  • ACQUITY Arc
  • 2998 Photodiode Array Detector
  • ACN
  • H2O

CAUSE

Poor mixing

FIX or WORKAROUND

Change the plunger stroke volume from 132 µL (default) to 66 µL.
With a 66-µL stroke volume: 
- This does not change the dwell volume
- It is useful for long, shallow gradients
- It reduces the baseline noise caused by pump pressure fluctuations 
- The maximum flow rate that the Arc Solvent Manager delivers will be limited to 2.5 mL/min and below

ADDITIONAL INFORMATION

Depending on the method, the path 1 mixer may not exhibit the cycling baseline symptom because it is longer and provides better mixing. 

Not able to find a solution? Click here to request help.