Random dips in baseline for 2414 RID - WKB69281
Article number: 69281
SYMPTOMS
- Intermittent dips in baseline
ENVIRONMENT
- 2414 Refractive Index Detector
- Empower 3 Enterprise
- Empower 3 Personal
CAUSE
Air bubble in the flow cell
FIX or WORKAROUND
Purge both sides of the flow cell with methanol for 30 minutes.
ADDITIONAL INFORMATION