Unstable baseline in GPC analysis using chloroform with 2414 - WKB69173

Article number: 69173

SYMPTOMS

  • Unstable baseline

ENVIRONMENT

  • 2695 Separations Module
  • 2414 Refractive Index Detector
  • GPC analysis using chloroform as mobile phase

CAUSE

The temperature of the column and detector is set to 40 °C.

FIX or WORKAROUND

Change the temperature of the column and detector from 40 to 35 °C.

ADDITIONAL INFORMATION

 