Unstable baseline in GPC analysis using chloroform with 2414 - WKB69173
Article number: 69173
SYMPTOMS
- Unstable baseline
ENVIRONMENT
- 2695 Separations Module
- 2414 Refractive Index Detector
- GPC analysis using chloroform as mobile phase
CAUSE
The temperature of the column and detector is set to 40 °C.
FIX or WORKAROUND
Change the temperature of the column and detector from 40 to 35 °C.