ACQUITY UPLC I-Class fails dynamic leak test with end point set to column - WKB68419
Article number: 68419
SYMPTOMS
- Post-PM testing
- Passes dynamic leak test to vent valve
- Above acceptable limit in dynamic leak test with end point set to column
ENVIRONMENT
- ACQUITY UPLC I-Class
- ACQUITY UPLC BSM
- ACQUITY Premier BSM
CAUSE
The fitting between the vent valve and the filter leaks.
FIX or WORKAROUND
- Replace the tubing between the vent valve and the filter or between the vent valve and t-piece on the Premier
- Rerun the dynamic leak test.
ADDITIONAL INFORMATION
Valve port 5 to tee - solvent line B: part number 430001207
Valve port 2 to tee - solvent line A: part number 430001208
You can replace the mixer with a union to determine which line is failing.