How to read energy on PDA detector - WKB68243

Article number: 68243

OBJECTIVE or GOAL

Read energy on an ACQUITY PDA.

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • 2998 Photodiode Array Detector 
  • ACQUITY UPLC PDA eλ Detector

PROCEDURE

  1. Lamp should be ON and lit for 60 minutes prior to reading energy.
  2. Launch the ACQUITY Console.
    • For 2998, right click on the 2998 window in the Run Samples screen, and select "Launch Console".
  3. In the left-hand tree, click on the PDA.
  4. Select "Maintain" menu.
  5. Then "Read energy".
  6. Set "exposure time" to 5 msec.
  7. Check "Continuous read".
  8. Check "Open" for the shutter position.
  9. Select the wavelength closest to 230 nm.
  10. Click "Read".

ADDITIONAL INFORMATION

Run with 100% HPLC-grade methanol in the flow cell at 230 nm. Example:

clipboard_ea05f7da2534d5552053cc6f4568c2e6e.png


 