FLR Instrument Failure in Empower - WKB67582

Article number: 67582

SYMPTOMS

  • FLR Instrument Failure when trying to set up a method in Run Samples.
  • Other Ethernet controlled instruments show "Yes" in node properties.
  • FLR shows "No" in node properties.
  • Restarting the FLR did not resolve this.

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC FLR Detector

CAUSE

Unknown.

FIX or WORKAROUND

  1. Power-off all ACQUITY components.
  2. Restart the LAC/E box.
  3. Allow LAC/E box to completely initialize.
  4. Power-on all ACQUITY components and allow them to completely initialize.
  5. Open Node Properties and verify that all components now say "Yes".
  6. Open Run Samples and set up the method.

ADDITIONAL INFORMATION