A loud squeaking or grinding sound is heard from Alliance 2695 sample compartment when selecting a tray - WKB66835
Article number: 66835
SYMPTOMS
- Loud noise from the Alliance 2695 when selecting a sample tray
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Sample trays are scraping along the skid plate as the carousel rotates.
FIX or WORKAROUND
- Remove all the sample trays and rotate the carousel to confirm that the noise has stopped.
- Feel along the bottom rim of the sample trays to confirm that the edge is very rough.
- After confirmation in steps 1 and 2, replace the sample trays with new ones (WAT270328 - Carousel Set 5pk kit).